Health

FDA classifies esophageal protection device for cardiac ablation as Class II with special controls

United StatesFinal RuleMedium ImpactView source ↗

The FDA final order classifies the mechanical deviation device for esophageal protection during cardiac ablation procedures into Class II with special controls under 21 CFR 870.5710. The rule applies to devices placed in the esophagus to mechanically deviate it away from ablation energy and requires clinical, animal, and non-clinical performance testing, biocompatibility, sterility, shelf-life, and labeling controls. It also means the device remains subject to 510(k) premarket notification requirements unless FDA later exempts it under section 510(m). The order is effective September 10, 2026, and FDA states the classification was applicable on September 6, 2023, when the De Novo order issued to S4 Medical Corp. took effect.

AI-generated summary. It may contain errors and is not legal advice. For the authoritative text, see the official source.

Key Changes

  • FDA codified a new device classification at 21 CFR 870.5710 for the mechanical deviation device for esophageal protection during cardiac ablation procedures.
  • The device is classified as Class II with special controls rather than remaining automatically in Class III under 21 U.S.C. 360c(f)(1).
  • Manufacturers must support the device with clinical performance testing, animal performance testing, non-clinical performance testing, sterility data, biocompatibility data, shelf-life data, and required labeling.
See 2 more changes with Pro

Affected Parties

  • Medical device manufacturers
  • Cardiac ablation device developers
  • +3 more

Tags

FDA · Class II · De Novo · +3

Related regulations

United States