FDA classifies cooperative powered surgical assist devices for ENT surgery as Class II with special controls
This FDA final order adds 21 CFR 874.4460 and classifies the cooperative powered surgical assist device for ear, nose, and throat surgery as a Class II device with special controls. The order became effective on September 10, 2026, and FDA says the classification was applicable on July 19, 2023. Manufacturers must meet specific testing, software, sterility, biocompatibility, and labeling requirements, and the device remains subject to 510(k) premarket notification because FDA did not exempt it under section 510(m). The device type was identified in the rule as a system that works with the surgeon’s movements to maintain direct physical control while improving precise and stable instrument positioning.
AI-generated summary. It may contain errors and is not legal advice. For the authoritative text, see the official source.
Key Changes
- FDA added 21 CFR 874.4460 to Part 874 and created a new classification for cooperative powered surgical assist devices for ENT surgery.
- The device type is classified as Class II with special controls, rather than remaining in the default Class III category for postamendments devices.
- Manufacturers must support the device with simulated-use testing, non-clinical performance testing, software verification and validation, hazard analysis, electromagnetic compatibility, electrical safety, thermal safety, sterility, shelf-life, biocompatibility, and labeling data.
Affected Parties
- Medical device manufacturers of ENT surgical systems
- FDA-regulated device sponsors seeking De Novo or 510(k) pathways
- +3 more
Tags
FDA · medical devices · ENT surgery · +3
Related regulations
United States