FDA Class II Classification for Whole Room Microbial Reduction Devices
This FDA final rule classifies the whole room microbial reduction device as a Class II medical device under 21 CFR 880.6510 with special controls. It applies to devices used to reduce microbial load on medical device surfaces after cleaning and disinfection, and it requires performance testing, biocompatibility testing, software validation, electromagnetic compatibility, electrical safety, and specific labeling. The rule was effective September 10, 2026, and FDA states the classification was applicable on September 1, 2023. Manufacturers of these devices must still submit a premarket notification under section 510(k) unless FDA later exempts the type under section 510(m).
AI-generated summary. It may contain errors and is not legal advice. For the authoritative text, see the official source.
Key Changes
- FDA codified 21 CFR 880.6510 and classified whole room microbial reduction devices as Class II with special controls.
- Manufacturers must show non-clinical performance testing for microbial log reduction, worst-case simulated use, in-use performance, photobiological safety, unoccupied-environment operation, and long-term material compatibility.
- Manufacturers must provide biocompatibility data showing safe residual chemical levels on device surfaces and/or gaseous byproducts in air.
Affected Parties
- Manufacturers of whole room microbial reduction devices
- FDA-regulated medical device sponsors
- +3 more
Tags
FDA · Class II medical device · 21 CFR 880.6510 · +3