FDA class II classification for temperature regulation devices used to protect the esophagus during cardiac ablation
This FDA final order classifies the temperature regulation device for esophageal protection during cardiac ablation procedures into class II with special controls under 21 CFR 870.5720. The rule applies to devices placed in the esophageal lumen to reduce esophageal injury or other adverse events during cardiac ablation, and it requires clinical, animal, and non-clinical performance data, sterility evidence, biocompatibility, shelf-life support, and specific labeling. The device was the subject of a De Novo request received on March 30, 2023, and FDA issued the classification order on September 13, 2023; the codifying final order is effective September 10, 2026. Manufacturers of these devices must comply with the special controls and, because FDA has not exempted this device type under section 510(m), it remains subject to 510(k) premarket notification requirements.
AI-generated summary. It may contain errors and is not legal advice. For the authoritative text, see the official source.
Key Changes
- FDA added 21 CFR 870.5720 to Part 870 to codify the device type as a class II cardiovascular device with special controls.
- Manufacturers must demonstrate clinical performance, including reduction of esophageal injury, lack of harmful effects on the ablation procedure, and absence of esophageal or oral injury.
- Manufacturers must provide animal testing, non-clinical testing on mechanical integrity, temperature change rates, and accessory compatibility, plus sterility and biocompatibility evidence.
Affected Parties
- Medical device manufacturers developing esophageal protection systems for cardiac ablation
- De Novo applicants and 510(k) submitters in the cardiovascular device sector
- +3 more
Tags
FDA · medical device classification · De Novo · +3
Related regulations
United States