FDA Class II Classification for Vaginal Hydrogel Packing Systems
The FDA issued a final amendment and final order classifying the vaginal hydrogel packing system into class II with special controls under 21 CFR 892.5735. The device is a non-powered positioning device used to reduce radiation dose to adjacent pelvic organs during radiation therapy planning and delivery. This means manufacturers must comply with specific testing, biocompatibility, sterility, shelf-life, and labeling requirements, and the product remains subject to premarket notification under section 510(k) because FDA has not exempted it under section 510(m). The rule is effective September 10, 2026, and FDA states the classification was applicable on August 22, 2023.
AI-generated summary. It may contain errors and is not legal advice. For the authoritative text, see the official source.
Key Changes
- FDA added 21 CFR 892.5735 to classify vaginal hydrogel packing systems as Class II devices with special controls.
- Manufacturers must generate clinical performance data showing the device works as intended, including radiation dose to adjacent organs, device stability, deployment and removal, and patient comfort.
- Manufacturers must conduct non-clinical performance testing, including radiation-integrity testing, bioburden testing if not supplied sterile, and structural integrity testing for tensile strength, leakage, and burst strength.
Affected Parties
- Medical device manufacturers making vaginal hydrogel packing systems
- Radiation oncology device developers
- +3 more
Tags
FDA · medical devices · radiology · +3
Related regulations
United States